RANA-Medical -- Patient First Respiratory ServicesRANA-Medical is a CCHSA-Accredited Company (opens in a new window)
:: Home :: Our Mission :: Products :: Respiratory Services :: Sleep Services :: Hospital Equipment :: Mine Safety ::
:: Contact Us ::
Privacy Policy :: Links ::
Hospital Oxygen Concentrators, Medical Gas Alarms, Medical Vacuum Systems
  Home > Hospital Equipment > Medical Gas Alarms
 


Advanced Design Combines Safety and Reliability

The RANA-Med Medical Gas Alarm Panel has been designed using the latest in solid state electronics, bringing to the hospital a monitoring system which provides the basics in safety, operating efficiency, and reliability.

As a fully electronic system, the RANA-Med Alarm Panel requires little regular maintenance. In accordance with the latest CSA Medical Gas Standards (Z305.1 and Z305.6-1992), the system is completely "self-supervised" with alarm functions occurring on an open circuit. Thus, wire breaks and faults will also be reported immediately. The RANA-Med Alarm Panel also incorporates a built-in battery back-up. Disruption of power will cause an alarm, while still permitting normal functions to occur.

All alarms are audible and visual, with each panel capable of reporting six different conditions.


Easy to Install - New or Renovated Facilities

Installing the panel is very straight forward, requiring only a wall to mount it on and a 120 VAC outlet.

It is suitable for new or renovated medical facilities.


Total Flexibility - Use With Any Medical Gas Supply

Whether it's Oxygen, Nitrous Oxide, Medical Vacuum, or Medical Air, monitor your medical gas delivery system with the increased safety, efficiency and reliability of a RANA-Med Medical Gas Alarm Panel.

Any of the following alarm conditions can be incorporated into one or more RANA-Med Panels:

Alarm Condition
Oxygen
Nitrous Oxide
Med Vac
Med Air
Abnormal Line Pres High
x
x
x
x
Abnormal Line Pres Low
x
x
x
x
Reserve Supply Low
x
x
-
x
Reserve Supply In Use
x
x
-
x
Secondary Supply Low
x
x
-
-
Secondary Supply In Use
x
x
-
-
Low Liquid Level
x
x
-
-
Low Head Pressure
x
x
-
-
High Air Temperature
-
-
x
x
Lag Alarm
-
-
x
x
Oxygen Concentrator Malfunction
x
-
-
-
Others, as Required
x
x
x
x

In accordance with Rimer Alco policy of continued development, specifications may change without notice.

 


205 Stephen Street
Morden, MB R6M 1V2
(204) 822-6595
rana@ranamedical.com


Copyright © 2004-2008 Rimer Alco North America. All Rights Reserved.

:: Site Map ::